By Rick Rechen
The following discussion of
Picosystem® accessories
builds on previous
presentations. There isn’t
sufficient time to repeat
all of the explanations, so
if something seems unclear,
refer to earlier
publications of Water Works.
The Picotech® 2 outlet is
another way to answer the
question about getting
sterile water from a Pico
System, posed by the user in
last months Water Works.
HOW A PICOTECH WORKS
The Picotech® 2 outlet
contains an Ultra Violet
(UV) Sterilizer with a low
ozone (254 nano meter (nm)
wavelength) UV lamp. It
exposes water from the
Picosystem® to high
intensity UV radiation. UV
radiation at the 254 nm
wavelength is very effective
at killing bacteria (bugs)
and other microorganisms.
The way it does this is to
break up the DNA of the
microorganism. When a
microorganism with broken,
non-working, DNA tries to
reproduce itself and
multiply, it can’t. So, when
it dies that’s the end of
it.
Since the UV light is the
last thing that the water
sees before it exits the
spout, the water should have
no living microorganism in
it and should be sterile.
The only thing that can add
back bacteria to the water
is contamination from the
Picotech® 2 spout. For this
reason, the nut connecting
the spout to the UV unit
must be very tight to keep
air from being drawn in.
Also, no one should touch
the end of the spout. That
could easily leave bugs
growing there.
It’s very important that the
correct lamp be used in the
Picotech® 2. There are two
kinds of lamps. The 254 nm
(low ozone) lamp is
effective at killing
bacteria. This is the one
used in the Picotech® 2. The
other lamp, the 185 nm (high
ozone) lamp, is good at
breaking up organic
substances, however carbon
dioxide gas, produced from
the broken up organics
lowers the specific
resistance of the water from
18 megohms to around 12
megohms. So, the 185 nm
lamp, used in the UV Plus
Pump Assembly, must always
have a deionizer after it.
WHO MIGHT BENEFIT
FROM A PICOTECH 2?
A Picotech® 2 outlet could
be provided for a customer
intending to use water for
preparing buffers or
reagents that could be
degraded by bacteria.
Someone filling incubators
where bacterial growth in
the water chamber is a
problem might also
appreciate a source of
bacteria free deionized
water. Just about any
general-purpose application
where microorganisms are of
concern could be considered.
Typical ones are Biochemical
Oxygen Demand and Clinical
Chemistry. Please note:
Even though water from a
Picotech® 2 might have no
living bacteria in it, there
still might be endotoxin.
SPECIFICATIONS FOR WATER
FROM A PICOSYSTEM® WITH A
PICOTECH® 2
Resistance: At Least 18 meg
TOC: <35 ppb
CFU: <0.10 CFU/ml
WHAT DOES
RECIRCULATION DO?
Adding recirculation to a
Picosystem® with a Picotech®
2 reduces numbers of bugs
growing in the system and
level of TOC in the system.
But, since water is not
recirculated through the
Picotech® 2, it doesn’t
really do anything to affect
its performance. A
Picosystem® Plus with a
Picotech® 2 will operate
with the same specifications
except that TOC will be less
than 10 ppb.
WHAT ABOUT ADDING
185 NM WAVELENGTH UV TO THE
PUMP?
Again, the Picotech 2
attaches to the end of the
Picosystem® so that
recirculated water doesn’t
pass through it. Water is
pumped through the 185 nm UV
unit in the pump assembly
and through the Picosystem®
over and over. Multiple
passes through 185 nm (high
ozone) UV reduce TOC from
less than 10 ppb to less
than 5 ppb.
WHO MIGHT WANT A
COMBINATION OF PICOTECH® 2
AND PLUS PUMP OR UV PLUS
PUMP?
Any application that
requires chemically pure
water and is sensitive to
living bacteria in the water
might benefit from this.
Specifically, many of the
same users who would like a
Picosystem Plus or a
Picosystem UV Plus might
want to add a Picotech 2.
These include Amino Acid
Analysis, GC-MS, DNA
sequencing and automated
clinical chemistry.
By Rick Rechen
Microorganism control is the
primary reason for using
ultraviolet and
recirculation in purified
water systems. A second
reason is to reduce TOC
levels.
Minimizing microorganism
levels in purified water
systems involves two related
concepts: Prevention and
Control.
In order to produce a
sterile environment,
specific efforts must be
carried out, or organics
will be present.
ULTRAVIOLET RADIATION
UV (254 nm) acts as a
barrier to living organisms
carried by a moving stream
of water. It does not effect
microorganisms attached to
piping upstream or
downstream of the UV unit.
It has been theorized that
for every microorganism
floating in the water
(killed by UV) there are
10,000 on the surface of the
piping (unaffected by UV).
At the very point of use (Picotech®,
Picopure®), 254 nm UV can
produce a sterile product.
However, it is more
difficult to determine what
245 nm UV achieves
“in-line.” The amount, or
activity of, endotoxin is
not reduced by 254 nm UV.
Exposure to high intensity
185 nm UV, over a period of
time, will break down
endotoxin and other
microorganism by-products,
and reduce TOC levels.
RECIRCULATION
Recirculation, as part of a
synergistic design, promotes
lowered microorganism
counts. A synergistic design
is a system design where the
combined action of 2 or more
pieces of equipment achieve
an even greater effect than
that of which each piece of
equipment is individually
capable. Some of the
synergistic pieces that make
recirculation effective are:
Smooth Piping and Joints.
Smooth interior piping
surfaces and joints make it
more difficult for
microorganisms to adhere to
the piping and allow fewer
places for microorganisms to
hide. Smooth surfaces and
crevice free joints allow
disinfectant to reach all
parts of the system, so that
it does a more complete job.
No Dead Legs. A
dead leg is defined as an
un-recirculated section of
piping longer than six times
the pipes diameter. That is,
a dead leg in a 1/2”
diameter piping system would
be any non-recirculated
section of piping more than
3” long. Dead legs create
all sorts of problems in a
piping system. The most
common problem is allowing
an area where microorganisms
and endotoxin accumulate.
Closed Loop. It is
preferable that once highly
purified water enters the
distribution system, it does
not come in contact with air
again. Air recontaminates
the water with carbon
dioxide and other gasses.
It has been found that most
bacteria do not survive and
reproduce well in highly
purified water. As long as
the deionizers are producing
water with high resistance
(is ion free), CFU counts
are low. Once ion
breakthrough begins to
occur, high CFU counts and
endotoxin levels accompany
the ions. In a closed-loop
you maintain the
micro-environment of high
resistance with low CFU
levels. When you return DI
water to a storage tank,
with higher nutrient RO
water, CFU counts and
endotoxin levels increase.
Storage facilities should be
sealed, with incoming air
being filtered through a
pre-sterilized hydrophobic
0.2 micron vent filter.
Regular disinfection of the
storage tank, or chlorine
injection, is a good idea to
keep microorganisms and
their by-products from
accumulating.
Also, remember that opening
storage tanks, sticking
hands in the stored water,
pulling out and replacing
level controls are all
potential sources for
microbial contamination in
storage and downstream
equipment.
SUMMARY
In-line 185 nm UV in a
recirculated system will
reduce overall TOC and, more
specifically, will break
down endotoxin, proteins,
enzymes and other
microorganism by-products.
Application of 254 nm UV at
points of use, such as a
Picotech 2 or Picopure 2
systems, can produce sterile
water.
Continuous recirculation of
purified water in a
distribution system designed
without dead legs and other
sources of microbial
contamination helps to
provide effective control of
microorganism growth. On a
small scale, the Hydro 17008
pump or UV Plus Pump will
perform this task.
By Keith Shepard
ORGANICS AND MICROORGANISMS
ARE DIFFERENT
The category of water-borne
impurities known as Organics
includes all carbon
containing compounds.
Bacteria and other
microorganisms contain
carbon, and are alive,
therefore they have their
own separate category.
ORGANIC REMOVAL
Filtration: Some
organics, such as fragments
of plant life, are not
dissolved but are in a
particle form. These can be
removed by filtration. The
level of filtration may be:
screens, depth media,
microfiltration,
ultrafiltration or reverse
osmosis.
Activated Carbon:
The microporous structure
within carbon attracts and
holds organics by Van der
Walls force. Organic
adsorption by activated
carbon is an effective means
for removing a broad range
of organics from water.
Recirculated deionizer
systems, using activated
carbon, can produce water
with TOC levels from 5-20
ppb.
Ultraviolet radiation:
With sufficient intensity
and exposure time, most
organic bonds between carbon
and other elements can be
broken with 185 nm UV.
Longer exposure times (3-4
times sterilization time),
or repeated exposure (such
as recirculation), is
necessary to achieve low TOC
levels. Recirculated
deionizer systems using both
activated carbon and 185 nm
ultraviolet can produce
water with TOC levels of 1-5
ppb.
CONTROL OF ORGANICS IN
PURIFIED WATER SYSTEMS
Here are some things we can
do to help control organics
in purified water systems:
1) Avoid using things which
add organics. a) Use aged
resins - New resin contains
a lot of organics as a
result of the manufacturing
process. b) Pre-rinse RO
membranes - new membranes
contain a surfactant
(organic) that must be
rinsed off.
2) Protect the system from
the intrusion of organics.
a) Good pre-filters to
remove organic particles. b)
Vent filters on storage
tanks.
3) Design the system to
protect itself. a) Use
continuous recirculation. b)
No dead legs. Dead legs are
a breeding ground for
microorganisms, which
increase the organic content
of the purified water. c)
Use 185 nm UV in a
recirculated distribution
loop. Multiple passes
through the 185 nm UV breaks
down many organic substances
by photo-oxidation.
By
Keith Shepard
INTRODUCTION:
Before discussing how a
Picosystem® works, lets
discuss what a Picosystem®
works on.
Hydro groups contaminates
into six general categories:
dissolved solids, organics,
particulates,
microorganisms,
microorganism by-products
and dissolved gases.
A typical 11218 Picosystem®
consists of a pr- filter
housing with a 0.2 micron
filter cartridge, activated
carbon canister and two
mixed-bed deionizers. It is
designed to purify water by
means of a series of
treatments, each with a
fairly specific purpose.
Which contaminant does each
treatment step address?
PREFILTER
The pre-filter is a critical
part of a Picosystem®.
Generally, the better the
pre-filter, the better the
Picosystem® output quality.
A pre-filter can also be a
source of contamination.
Some membrane pre-filters
have a wetting agent
(surfactant) which coats the
surface. This wetting agent
must be rinsed off, by
flushing the filter.
Surfactant not rinsed from
the pre-filter will pass
through the Picosystem® and
can interfere with some
applications.
NOTE: The
Polyethersulfone filters
that we now get from Parker
are naturally hydrophilic.
Hydrophilic filters do not
require a wetting agent, so
there is none to rinse off
from the Parker
Polyethersulfone filters.
ACTIVATED CARBON
Granular activated carbon is
employed to remove chlorine,
chloramines and dissolved
organics. In doing so, it
provides a chlorine-free
environment with a lot of
nutrient for microorganism
growth. Hydro recommends the
replacement of carbon tanks
every six months to prevent
accumulated microorganism
and microorganism
by-products from
overwhelming the rest of the
system.
Carbon fines, generated
during shipping can also be
a problem. If the carbon
tank is not flushed to
remove these fines, when the
tank is installed carbon
particulates can pass
through the system and
contaminate the final
product.
MIXED-BED RESINS
New resin contains organics
from the manufacturing
process. If not removed by
several regeneration and
exhaustion cycles, which
expand and contract the
beads (like squeezing a
sponge), they can
contaminate the product
water.
NOW, HOW DOES A
PICOSYSTEM® WORK?
Deionization is a process
involving the removal of
dissolved mineral salts. The
dissolved salts are
preferentially exchanged for
other ions that are already
on the resin bead. The ions
that are already on the
resin beads (from our
regeneration) are: Hydrogen
(H+) on cation beads and
hydroxyl (OH-) on the anion
beads. So, mixed-bed resin,
with Type I anion resin, is
capable of removing all
ionized species and
producing 18+ Megohm H2O.
CONCLUSION
Tap water contains all sorts
of particles, organics,
microorganisms, dissolved
salts, gasses, etc. The
pre-filter removes
particulates, activated
carbon treats
chlorine/chloramines and
dissolved organics,
deionizers remove dissolved
solids (salts) and carbon
dioxide. Each treatment step
is fairly specific in what
it removes and one cannot
compensate for another’s
failure.
By Keith Shepard
INTRODUCTION:
We will now discuss
equipment that can be added
after the Picosystem®, to
make the water even better.
In general, we can add
ultrafiltration, ultraviolet
(254 nm and 185 nm) and
recirculation. Sometimes we
add a combination of these
three things.
Purified water from a
standard Picosystem® will
have the following
specifications:
Resistance - >18 megohm
TOC - <35 ppb
CFU - <10 CFU/ml
Endotoxin - ??
WHAT CAN BE IMPROVED
WITH POST-TREATMENT
EQUIPMENT?
SPECIFIC RESISTANCE will not
be increased by any of the
three post treatment steps.
TOC can be improved with the
addition of recirculation.
Recirculation can be
provided by a Plus Pump or
UV Plus Pump. The Plus pump
will lower TOC levels to <10
ppb. The UV Plus pump (with
185 nm UV) will lower TOC
levels to <5 ppb.
To reduce bacteria (CFU),
the least expensive way is
to add a Picotech® 2 unit,
which contains 254 nm UV.
The CFU’s will be reduced to
<0.1 CFU/ml.
To reduce endotoxin requires
the addition of an
ultrafilter. The Picopure®
2, with an ultrafilter and
254 nm UV will reduce the
endotoxin level to <0.25 eu/ml.
This unit will also reduce
the CFU’s to <0.1 CFU/ml.
This is a combination unit
since it contains both UV
and Ultrafiltration.
Other combination units
would be the Picotech® 2 UV
Plus. This unit will reduce
both TOC and CFU’s.
A Picopure® 2 UV Plus will
lower TOC, CFU’s and
endotoxin.
SUMMARY:
Our standard Picosystem® can
perform even better with the
addition of some
accessories. Just determine
each customers needs, and
then provide the proper
system/post treatment
equipment to meet those
needs.
By Keith Shepard
INTRODUCTION:
Many materials have been
used to transport purified
water. Things like glass,
stainless steel, tin and all
kinds of plastic. This
article will discuss
different plastic piping
materials.
TUBING/PIPING
Water has to be transported
from a Picosystem® to the
point of use in something.
Polyethylene and
Polypropylene have proven to
be popular choices. Black PE
tubing does not admit light.
This helps reduce the growth
of bacteria and algae.
Black PE tubing and Beta
Polypropylene pipe contain
ultraviolet inhibitors.
These UV inhibitors protect
them from becoming brittle
due to exposure to
ultraviolet light. The
ultraviolet light might be
from sunlight or fluorescent
lights.
Numerous other types of
tubing exist. There is
tubing made of Nylon,
Polypropylene, Teflon and
PVC. One type of tubing that
we often see in a laboratory
is Tygon tubing. Tygon is
made of PVC. By itself, PVC
is not flexible. To make it
into flexible tubing,
organic compounds called
plasticisers are added.
These organic plasticisers
leach from the tubing
throughout its life and can
interfere with many types of
research, especially HPLC or
GC-MS. Also, bacteria love
to eat the plasticisers in
Tygon.
It should be noted that new
tubing/piping is not
sterile. Disinfection of a
new tubing/piping system may
be required, depending on
the application.
FAUCET OUTLET
How to best dispense water
from a Picosystem® is a good
question. Hydro is one of
the few ultrapure water
companies that makes faucet
assemblies.
Our Picotaps® are
constructed from UHMW PE.
Picotaps® contain a
diaphragm valve and will
last for many years.
Diaphragm valves are
generally cleaner than other
types of valves and
replacement of the diaphragm
is a simple, inexpensive
procedure.
TUBING ON FAUCET IS
BAD
Many users attach a section
of tubing onto the outlet,
either to fill large
containers or reduce
splashing in the sink. Hydro
does not recommend this, but
instead recommends the use
of a hand held faucet or a
Flex-tap® faucet.
Nonetheless, tubing is
sometimes attached to
outlets. Rubber is the
worst. All sorts of
microorganisms like to eat
rubber. Sometimes tubing is
long enough to lie in the
sink. This enables tubing to
be contaminated not only
with microorganisms from the
air and the Picosystem®, but
also from anything that is
dumped into the sink.
WHAT DOES IT DO TO ADD A
PUMP TO A TYPICAL 11218
PICOSYSTEM AND WHY WOULD YOU
WANT TO DO IT?
By Rick Rechen
As simple as it is, the
basic Hydro Picosystem®,
consisting of
pre-filtration, organic
adsorption, two stages of
mixed-bed ion exchange, QA
light and a Picotap®
represents the limits of the
technologies it employs.
There just isn’t much more
you can do to improve it,
without changing one or more
of the treatment steps or
adding something to the
system.
One thing that can be added
to a Picosystem® is a Plus
Pump to recirculate the
water.
What would this do?
It does several things, but
it may make more sense to
describe the changes that
recirculation brings about
if we first talk about what
a Picosystem® does without
recirculation.
IMPURITIES THAT
PICOSYSTEMS REMOVE
There are six broad
categories of impurities in
most tap water. These are:
Particulates (dirt, sand,
silt etc…), Organics (Carbon
containing compounds),
Dissolved Solids (salts or
ions), Microorganisms
(bacteria, virus, algae,
fungus, mold, etc…),
Microorganism by-products (endotoxin,
proteins, ammonia, DNA, RNA,
RNase, etc…) and Dissolved
Gasses (mainly carbon
dioxide but also ammonia and
oxygen). Picosystems®® are
designed to go a long way
toward removing four
categories of impurities and
to keep the other two at
very low concentrations.
Water from a typical 11218
Picosystem®, that is used
and serviced regularly, has
certain overall
characteristics.
Particulates: It
has very few particles (from
using a 0.2 micron
pre-filter),
Organics: It has a
low level of organics (less
than 35 ppb TOC ),
Dissolved Solids:
It is essentially ion free
(at least 18 megohms
specific resistance)
Microorganisms: It
has a low number of bacteria
(<10 CFU/ml) and other
microorganisms in the water.
(This can vary with use
levels.)
Microorganism
by-products: Although
Microorganism by-products
are not covered by
Picosystem® specifications,
water from a Picosystem® is
typically endotoxin-free and
RNase-free (but, it may not
be).
Dissolved gasses: The
anion resin removes carbon
dioxide and carbon removes
dissolved ammonia so that it
is free from dissolved
gasses.
WHY USE RECIRCULATION
Limitations of the
Picosystem® relate to what
happens when water isn’t
being used. In fact, water
only moves when someone is
actually running it out of
the Picotap®. So, most of
the time water isn’t moving.
When water isn’t moving
microorganisms tend to
multiply in it. Multiplying
microorganisms tend to
produce endotoxin and other
by-products. So, straight
pass Picosystem® output is
more likely to have higher
bacteria counts and
endotoxin as well as DNA,
proteins and enzymes than
water that is kept moving.
In addition, water that
passes through carbon one
time and sits still tends to
have more organics than
water that passes through
carbon several times and
keeps moving. So, water from
a Picosystem® tends to have
more organics (higher TOC
readings) than water that is
recirculated. The maximum
TOC concentration specified
for water from a Picosystem®
is 35 ppb. Water from a
recirculated Picosystem® (Picosystem®
Plus) usually produces TOC
readings of less than 10
ppb.
ADDING 185 NM
ULTRAVIOLET TO RECIRCULATION
Including 185 nm (high
ozone) ultraviolet in the
Plus Pump assembly (UV Plus
Pump) is not the same as
adding 254 nm (low ozone)
ultraviolet to the end of a
Picosystem®. It doesn’t do
much more to reduce
microorganism levels than
the Plus Pump by itself, but
it does reduce organic
levels even further. Water
from a recirculated
Picosystem®® with a 185 nm
UV unit (Picosystem® UV
Plus) usually produces TOC
readings of less than 5 ppb.
WHEN TO ADD RECIRCULATION
WITH 185 NM ULTRAVIOLET
Most Hydro customers use the
water for applications where
the water doesn’t have to be
sterile but should have a
low number of bacteria. All
of these users would benefit
from recirculation.
Others are not concerned
with bacteria, but would
like to be sure that
organics remain low. Typical
applications: High
Performance Liquid
Chromatography (HPLC), Ion
Chromatography, Gas
Chromatograph-Mass
Spectrometry (GC-MS), Amino
Acid analysis).
Adding a UV Plus Pump to a
Picosystem® could meet that
requirement. To get the full
effect for microbial control
it is necessary to have a
fully recirculated Picotap
so that there is no water
standing anywhere in the
system.
MOST USERS WOULD BENEFIT
FROM RECIRCULATION
Most Hydro customers could
benefit from adding
recirculation to a straight
pass Picosystem® (to make a
Picosystem® Plus) and a
recirculated Picotap® faucet
because water quality would
improve and be more stable
than it is with a single
pass Picosystem® alone.
WHY ARE PICOSYSTEMS & QA’S
SO EXPENSIVE?
By Keith Shepard
INTRODUCTION:
The short answer might be
because of the quality
materials we use and the
excellent service we
provide. But, a better
answer might be, expensive
in comparison to what?
A Hydro Picosystem® normally
produces purified water at
about 1/3 the cost of
prepared bottled water,
water from a cartridge type
system or distillation. Here
in Durham, NC, the average
Picosystem®® will produce
about 340 gallons with tap
water feed. If the exchange
cost is $477.00, then the
cost per gallon of ultrapure
water is $1.40. This is for
purified water, on demand,
at 1/2 gpm with plain old
tap water feeding the
system. A cartridge system
has to have pretreated feed
and bottled water is not “on
demand.” The same arguments
apply to distilled water.
If the comparison is to
equipment from other
companies (that have tanks
and filter housings that
look like ours) our price
may be higher, for what
seems to be equivalent
equipment. Now is the time
to discuss the quality of
material we use and the way
we process it.
CARBON:
Hydro selects activated
carbon that is manufactured
from a metallurgical grade
of bituminous coal with low
ash content and a high
degree of abrasion
resistance. The activated
carbon has a good
distribution of large and
small pore sizes to remove
the broad spectrum of
organics from water. Many
competitors use carbon that
is primarily designed for
removal of chlorine and
taste and odor compounds.
Hydro selects activated
carbon based on its ability
to remove TOC in water.
New carbon has a lot of
fines to be flushed from it.
Before you can flush the
fines from new carbon it has
to be hydrated (soaked with
water). This hydration
process can take several
hours, but it is an
important step in preparing
carbon to go into our tanks.
Our tanks are cleaned and
sanitized before carbon is
put into them.
RESIN:
The primary consideration in
the selection of resin is
its ability to produce the
best possible water quality.
This not only includes the
ability of mixed resin to
produce 18 megohm water, but
the integrity of the resin
to prevent degradation of
the water by introducing
organic extractables and
particulates.
Cation resin is made both as
strong acid and weak acid
cation resin. For
demineralization, strong
acid cation resin is almost
always used.
Anion resin is made as both
strong base and weak base
resin. Strong base resin is
also made as Type I and Type
II. Some competitors prefer
to use Type II strong base
anion resins because they
cost less and can be
regenerated less
expensively. Type II resins
are less resistant to high
temperature, oxidation, and
physical attrition. Type I,
strong base, anion resin is
required for the production
of ultra high purity water.
Only strong base, Type I,
anion resin will remove
weakly ionized carbon
dioxide and silica. This is
why Hydro uses strong base
Type I anion resin, even
though some of the
associated costs may be
higher.
Why isn’t the pH of my
purified water 7?
By Rick Rechen
INTRODUCTION:
“Why doesn’t the pH read 7.0
from the purified water
system?
WHERE DO WE GET PH?
The term, pH,is a
mathematical expression for
the concentration of
Hydrogen ions in solution,
and is commonly used to
describe how acidic or basic
water is. A pH of less than
7 indicates that there are
more Hydrogen ions than
Hydroxide ions and is
acidic. A pH greater than 7
indicates that there are
more Hydroxide ions than
Hydrogen ions and is basic.
A pH of 7 indicates that
there are equal numbers of
Hydrogen and Hydroxide ions
and is neutral.
RESISTANCE OF DI
WATER
Water by itself does not
conduct electricity very
well and has a high
resistance to electrical
flow. One way in which the
quality of deionized water
is expressed is by its
Specific Resistance. Good
quality deionized water has
a specific resistance of at
least 10 million ohms (megohm).
The more ions present in
water, the lower the
specific resistance. This
idea figures into the
measurement of pH in D I
water.
MEASUREMENT OF pH
Measurement of pH, by means
of a pH meter, involves two
electrodes. One electrode is
a reference electrode which
establishes the ionic
strength of the solution. It
does so under the assumption
that Chloride ion (Cl-)
represents the bulk of the
ions present in the solution
and uses a Chloride (also
known as Calomel) electrode
to establish the reference
point. The second electrode,
a Hydrogen ion electrode,
measures the concentration
of Hydrogen ion. The
electrical potential
difference between the two
electrodes is translated by
the meter into a number
called pH.
pH OF DI WATER
There are two problems in
directly measuring the pH of
deionized water.
THE FIRST
PROBLEM-WHY IS THE PH SO
ACIDIC (OR BASIC)?
There are no Chloride ions
in deionized water greater
than 10 megohms. So, the pH
meter reference electrode is
unable to set up an accurate
reference point from which
to measure pH. Therefore,
whenever pH meter electrodes
are placed, in freshly drawn
deionized water, the meter
provides a reading which is
incorrect. The reading
obtained will depend upon
the type of electrode
connected to the meter. Dual
electrodes usually give a pH
reading of around 5.8. A
combination electrode
usually gives a reading of
8.5. They will do this
simultaneously in the same
beaker of water.
To measure the pH of
deionized water it is
necessary to add a crystal
or two of Potassium Chloride
to the water. After this has
been done, pH will be around
6.5-6.8.
THE SECOND
PROBLEM-WHY ISN’T THE pH
7.0?
Just as soon as deionized
water is placed in a beaker,
it starts to absorb Carbon
Dioxide (CO2) from the
atmosphere. Carbon Dioxide
dissolves in water as
Carbonic Acid (H2CO3). The
small amount of Carbonic
Acid from dissolved Carbon
Dioxide will cause the pH to
shift from 7.0 to 6.5-6.8.
HOW CAN THE USER
KNOW THAT THE pH OF D I
WATER IS 7.0?
With typical laboratory pH
meters, in open beakers, it
is not possible to obtain a
pH of 7.0 in deionized
water. Adding Chloride ion
to the water will give only
an indication that the pH
might have been 7.0 before
the water was placed in the
beaker. However, it is
possible to answer the
question of pH indirectly.
Deionized water (>10 megohm
specific resistance) is
essentially ion free. (If it
were not, because dissolved
ions conduct electricity,
the resistance would be less
than 10 megohm.) If the
specific resistance is found
to be greater than 10 megohm,
then there are not enough
ions present to conduct
electricity. So, there is
also no excessive
concentration of Hydrogen or
Hydroxide ions. Therefore,
the pH must be 7.0 or
neutral. Conversely, if
there were enough excess
Hydrogen or Hydroxide ions
to shift pH to the acid or
base sides of the scale, the
specific resistance would be
less than 10 megohm. This is
the best we can do, at this
time, to answer this
question.
Why did I get only half the
normal amount
from my DI system?
By Keith Shepard
Let’s look at some things
that effect deionizer
capacity.
Deionizer capacity is a
function of three things:
1) Feed water quality
2) Feed water flow-rate
3) Regeneration state of the
resins
Feed water quality is
constantly changing. There
are two types of “tap
water”, water from deep
wells and surface water from
rivers and reservoirs. Well
water is usually more
constant in quality than
surface waters. It contains
less organics and more
inorganics. Surface waters
vary considerably from place
to place, time of the year,
sometimes from hour to hour.
Basing deionizer life on one
water sample is not ideal.
When deionizer life is in
question, we recommend
taking water samples weekly
until the cause is
determined.
Another problem associated
with feed-water is organic
acids. Organic acids can
blind the exchange sites on
resin beads and reduce the
quality and quantity that a
deionizer produces. We
cannot test for organic
acids in the lab but you can
test a tank in the field.
Remove a suspect tank and
connecting it to a water
supply where you are not
having a problem. If the
quality comes back up,
organic acids are probably
the problem. If organic
acids are a problem, the
usual treatment is to add
ultrafiltration before the
deionizers.
Flow-rate can
affect deionizer capacity.
Deionizers have both minimum
and maximum flow-rates. The
maximum flow-rate is
about four gpm per cubic
foot of resin. A flow rate
faster than this may not
allow time for complete
deionization. A fast flow
rate will show-up as ionic
break-through, or
exhaustion, before all
exchange sites on the resin
beads are used-up. This
translates into less than
expected throughput from the
deionizer.
The minimum flow-rate
is ½ gpm per square foot of
resin. This would be 1/10
(.1) gpm for 18” deionizers.
Slower flow will “channel”
through the resin bed, so
the entire resin bed does
not get used. The tank
appears exhausted, but only
about half of the resin in
the tank has been used. Is
your first thought, when you
have a customer with an
“early exhaustion” problem,
to put in a Kent water
meter? This can be a mistake
because the ¾” Kent water
meter (the only one
available until recently) is
not accurate below ¼ gpm.
The meter will not read
water flowing very slowly.
In the customers mind this
would confirm his suspicion,
that the DI tank only gave
half of its rated capacity.
We now have a ½” Kent water
meter (P/N 27250) that is
accurate down to 0.125 gpm
(1/8 gpm). This meter may
help us to identify low flow
problems, but it is possible
that a slow leak would still
not be detected.
The first thing to do in a
suspected low flow problem
is to look for a slow leak
somewhere in the system, a
very small solenoid valve,
or faucet that is left open
to flow really slow etc. One
fix for this problem may be
to install a recirculation
pump so that the flow rate
through the resin tanks is
always above the minimum
required flow rate.
The regeneration state
can also effect deionizer
capacity. For example, lets
look at the regeneration of
a water softener. A water
softener has only cation
resin and is generally
regenerated with salt
(sodium chloride). A water
softener has a capacity of
30,000 grains per cubic foot
of resin, when regenerated
with 15 pounds of salt per
cubic foot of resin. The
same softener has only
20,000 grains capacity per
cubic foot, when regenerated
with 5 pounds of salt per
cubic foot. Now, if you had
tap water with 8 grains per
gallon hardness the first
softener could treat 3750
gallons of water before
exhaustion. The second
softener could treat only
2500 gallons of water before
exhaustion. They both have
one cubic foot of media, but
the first was regenerated
with more chemical,
therefore giving more water
before exhaustion.
A similar phenomenon could
happen in the regeneration
of mixed-bed resins. This is
why at Hydro we use a “cook
book” approach to
regeneration. Every batch of
resin is regenerated the
same way. The chemicals used
are constantly scrutinized
for quality and are
precisely measured into each
regeneration batch. Also,
resin from each batch is
tested for capacity, and
records are kept on file.
Each tank filled also has to
pass a quality test before
being shipped.
As you can see this last
item is the one that we, at
Hydro, have the most control
over. It is actually the
least likely cause of a
premature exhaustion
problem. You will usually
find that the problem is
related to feed-water
quality, flow-rate or tank
exchange procedures.
We have not talked about
tank exchange procedures
yet, but they could include
air in the tanks, wrong flow
direction, carbon and resin
tanks switched, a bad cell
light, exhausted polisher or
worker/polisher mix-up.
WHY DOES THE WATER FLOW SO
SLOWLY?
By Keith Shepard
INTRODUCTION:
This question can have
several answers. The system
may be brand-new, or
recently exchanged. The
system may have been
operating just fine for
several months but now the
flow rate has slowed, or
stopped. The system may be a
tap-fed Picosystem®, or it
may be on a pretreated feed
source. It may be a single
pass system, or recirculated,
with a Picotech®2 or
Picopure®2.
I will lump the reasons for
low flow under three
categories. The three
categories are: Supply,
Plugging and Mechanical.
SUPPLY
I would recommend that you
first check the water supply
to the Hydro system. If the
system is on tap water, it
is possible that there is no
(or low) tap water pressure.
A valve upstream may have
been closed. The water
supply may have been cut-off
to the whole building. A
well pump may have failed.
If the system has a
pretreated feed source,
there could be some of the
same problems, and some
other problems. The storage
tank may be out of water.
There may be a problem with
the distribution pump.
PLUGGING
If you have supply water,
then the next thing to check
for is the Hydro Picosystem®.
Some components may be
plugged. Plugged also
applies to crimped, in the
case of tubing. Start with
the first component and work
your way through the system
until you fInd what is
plugged, or where the tubing
is crimped. The most likely
item will be the pre-filter
or carbon tank. Tubing can
get crimped when someone in
the lab moves things around
near the Picosystem®. If the
system contains a
Picopure®2, the ultrafilter
could be plugged.
MECHANICAL
If you do not find anything
plugged or crimped, you may
have a mechanical problem.
Two or three possible
mechanical problems are: 1)
A problem with the pressure
regulator, 2) a problem with
the faucet, or 3) a problem
with the pump.
You should have found that
there was a problem with the
pressure regulator while
checking the filter for
plugging. The faucet valve
would be another thing to
check. It is possible that
something is under the
diaphragm or the faucet stem
is not working properly. The
knob may be slipping, or
threads may be stripped.
A third mechanical problem
could be with the
recirculation pump. If the
pump is before the system,
is it the correct size, is
it running? Is the pump
actually pumping? If the
pump is after the system, it
can draw water away from the
faucet so that there will be
low flow at the faucet. It
has been found that it is
usually better to have the
pump before the system.
WHY SHOULD I EXCHANGE THIS
SYSTEM?
By Keith Shepard
Why should I service my
Hydro system every six
months even though the light
is on, or I haven’t used
much water?
The essence of research is
to gather valid data from
scientific studies that can
be duplicated later (or in a
different lab); under the
same controlled conditions.
To duplicate the same
controlled conditions you
need the same strength,
quality and purity of
chemicals as in the original
experiment. In many cases,
the most used chemical is
purified water. The purified
water may be used to wash or
rinse the glassware used in
the experiment, as well as
to mix or dilute other
chemicals. It may also be
the main ingredient. This is
the reason that the
consistency of the purified
water is critical to
research.
Chemicals, of guaranteed
quality, are purchased in
fairly small quantities, are
expensive and may have a
limited shelf-life. Water
from a Hydro Picosystem® is
no different. It is a
chemical that comes with
guaranteed consistency, is
fairly expensive and has a
limited shelf-life. Unlike
most other chemicals used in
the lab, the Hydro purified
water system can make water
on-demand, in fairly large
quantities, from either tap
water or pretreated water.
It also has a very
economical cost per gallon,
as compared to most other
chemicals, and it can be
used to make excellent
coffee.
How can we guarantee the
quality of the purified
water from a Hydro
Picosystem®? From over 34
years of experience we know
the quality that each type
of Hydro Picosystem® is
capable of producing. We
have verified these findings
with expensive and continued
testing. This testing and
experience has given us the
confidence to guarantee the
quality of our purified
water, if the six month
Quality Assurance Service is
performed. So, even if the
light is on, or they haven’t
used much water, the six
month Quality Assurance
Service is required for
Hydro to guarantee the
expected water quality.
UV Replacement Frequency
Explained
By Keith Shepard
HOW IT WORKS
Energy from a UV lamp
destroys the
deoxyribonucleic acid (DNA)
and prevents microorganisms
from reproducing. UV lamps
can be made to emit only 254
nm wavelength UV radiation,
or both 185 and 254. The 185
nm bulbs are used for TOC
reduction. The 185 nm
wavelength first promotes
the conversion of dissolved
oxygen to ozone. Ozone is
then changed by the 254 nm
wavelength UV light to
hydroxyl radicals. These
hydroxyl radicals then
oxidize organic molecules in
the water. With sufficient
contact time and/or
recirculation, the organics
are oxidized to CO2 and
water.
UV DOSAGE
Disinfection effectiveness
depends on the UV system
dosage. The UV dosage must
also take into consideration
the transmission qualities
of the source water. The
total intensity that the UV
lamp emits is measured in
microwatts at a specific
wavelength. By combining the
total lamp intensity and the
residence time, a UV dosage
can be derived. This dosage
is expressed in
microwatt-seconds per square
centimeter (mW-sec/cm2).
Most UV bulbs are
manufactured to be able to
provide a dosage of 30,000
mW-sec/cm2. This 30,000 mW-sec/cm2
is what the bulb is rated to
produce at the end of
8-9,000 hours (about a
year). This means that when
new, UV bulbs produce
approximately 60,000 mW-sec/cm2.
The reason that lamps are
rated in this manner is key
to understanding why we, at
Hydro, recommend a six month
replacement of the UV bulb,
and annual replacement of
the quartz jacket.
LAMP LIFE
Unlike a regular light bulb,
UV lamp performance
deteriorates with time.
Their life span is limited.
According to one expert,
they should be replaced
every 9 to 12 months for
maximum effectiveness. The
reason is scale build-up and
Solorization – a condition
in which the quartz
discolors. This
discoloration decreases the
intensity of the UV light.
Solorization affects both
the lamp material and the
quartz jacket. The effective
UV intensity decreases
approximately 40-60 percent
over 9,000 hours (just over
one year) of operation.
At this point, I would like
you to refer to Hydro
Technical Bulletin #1 “Is 18
Meg-ohm Good Enough”, and #3
“Quality Assurance: The Key
to Consistency”. If a
customer has a purified
water system, he/she wants
consistency from day to day.
To assure consistency, the
exchange frequency of some
items is time dependent
(Bulletin #3). The UV bulb
and jacket along with the
pre-filter and carbon tank
are such items. As is
pointed out in Bulletin #1,
there are six categories of
contaminants in tap water. A
standard Picosystem will
address four of these items.
A customer who’s Picosystem
contains a UV unit is,
obviously, trying to achieve
even better water quality.
His/Her system now addresses
five of the six contaminant
categories. It only seems
reasonable that our
recommendation on
maintenance should be
consistent with maintaining
the customer’s water system
in a manner that will not
allow a change in quality.
Only with a recommendation
of six month bulb and annual
quartz jacket replacement
can we feel confident in
achieving this goal.
Ion Exchange Explained
By Keith Shepard
PRINCIPLES OF DEIONIZATION
Deionization is the process
of removing dissolved,
ionizable solids from water
using the principles of ion
exchange. Ion exchange is a
reversible process. The
reversing of the process is
called regeneration. Ion
exchange works for two
reasons:
-
Resins prefer
“higher valence”
ions such as calcium
over lower valence
ions, like hydrogen
or carbon dioxide.
-
Resins are
influenced by the
mass of ions
present.
In a softener, the ion
exchange process essentially
exchanges the minerals
calcium and magnesium (also
iron and manganese) for
sodium. A water softener
requires only one type of
resin because it exchanges
only cations.
Mixed-Bed Deionization is
similar, but more
complicated because it
involves the removal of
cations and anions from the
water.
All dissolved minerals are
composed of both a metallic
part (a positively charged
cation) and a non-metallic
part (a negatively charged
anion). Mixed-Bed deionizers
require two resins because
they exchange both cations
and anions. No single resin
can exchange both cations
and anions. Ion exchange
depends on the tiny
electrical charges in which
like particles repel and
unlike particles attract
each other, like a magnet.
Cation exchange resin is
chemically formulated to
attract positive ions, and
anion exchange resin is
formulated to attract
negative ions.
MONO-BED
DEIONIZATION
Before progressing to
mixed-bed deionizers, lets
try to understand how a
mono-bed deionizer works. In
a two-column unit the cation
exchange resin is in one
pressure vessel and the
anion exchange resin in
another. Water first passes
through the cation tank,
then the anion tank. You
need to understand this
principle before learning
how a Mixed-Bed deionizer
functions. Here is what
happens when the cation
resin is in one tank and the
anion resin is in another
tank.
As water passes down through
the cation tank it
encounters millions of resin
beads, each of which contain
a large number of negatively
charged exchange sites in
the pores and microscopic
paths of its structure. When
the resin is in the
regenerated state, each
exchange site is occupied by
a positively charged
hydrogen ion (H+). As the
positively charged cations
in the water contact the
beads, they are attracted to
the negative exchange sites.
Since they have a stronger
positive charge than the
hydrogen ions they drive off
the hydrogen ions and attach
to the exchange sites. The
displaced hydrogen ions pass
down through the resin bed
with the water.
Because the hydrogen ions
are acidic, water from the
cation tank is a stream of
dilute mineral acid.
At the same time anions,
such as sulfates, chlorides
and other elements pass
through the cation tank
unchanged. Now everything is
piped to the anion tank.
ANION EXCHANGE
PROCESS
The anion exchange process
is similar to the cation
exchange process. Strong
base anion resin is made of
beads which have positive
exchange sites, which in the
regenerated state are
occupied by negative
hydroxide ions (OH-). As the
negatively charged anions
contact the beads, the same
attraction-repulsion process
takes place and the negative
hydroxide ions are dislodged
and replaced by the stronger
negative anions.
The hydroxide ions (OH-)
pass down through the anion
resin and are discharged
from the tank. At the same
time, the hydrogen ions (H+)
from the cation tank have
passed through the anion
resin unchanged. They now
join with the hydroxide ions
to form HOH, or H2O, or
water.
MIXED-BED DEIONIZATION
If the cation-anion
(mono-bed) exchange process
could be repeated many
times, the efficiency of ion
exchange would improve
remarkably. Since no
exchange process is 100
percent efficient,
successive ion exchanges
would be an improvement in
purity with each cycle. This
is exactly what happens when
cation and anion resins are
mixed together in a
Mixed-Bed Deionizer. As
water passes through the
mixed-bed deionizer it has
millions of chances to
contact a cation resin bead,
then an anion, then another
cation, another anion, and
so on. With each exchange,
the purity of the water
improves because more ions
are removed and held by the
resin beads. As water passes
down through the resin, the
cation resin will remove
cations, releasing hydrogen
into the water, and the
anion resin will remove
anions, releasing hydroxide
into the water. Released
hydrogen ions will ionically
bond with released hydroxide
ions to form a new water
molecule.
The Unique and Unusual
Properties of Water
Charles Riley, Jr.
PE
INTRODUCTION
To the average person, water
is a common and ordinary
substance which is often
taken for granted, that is
until a drought threatens
crops and drinking water
supplies, or a severe flood
destroys life and property.
Most people do not
understand that without
water, and its unique and
unusual properties, life, as
we know it on earth, would
not exist.
The abundance of water is
apparent. It is also the
only substance naturally
present on earth
simultaneously in three
distinct states or forms–
solid, liquid, and gas. On a
cold winter day, snow and
ice cover a field, while
water flows in a nearby
stream and gaseous clouds
float overhead.
FORMS OF MATTER
All substances exist in
three distinct forms (solid,
liquid, and gas). The form
of a substance, at a given
time, is a function of
temperature and pressure. A
solid is defined as matter
with rigidity and definite
shape, having a crystalline
internal structure. Solids
tend to resist external
forces. Solids can be
converted to liquids by
heating. The freezing point
temperature of pure water is
0° C, at one atmosphere
pressure. At temperatures
below the freezing point,
water exists as a solid -
ice.
A liquid, in contrast to a
solid, lacks rigidity and
has no definite shape. It
has a definite volume and
conforms to the shape of the
container in which it is
stored. External forces will
cause a liquid to flow.
Water is a liquid between
the freezing point
temperature (0° C) and the
boiling point temperature
(100° C), at one atmosphere
pressure. Liquids can be
converted to the gaseous
phase by heating beyond the
boiling point temperature.
A gas has neither rigidity
nor definite volume. A gas
conforms to the shape and
volume of the vessel
containing it. A gas greatly
expands and contracts with
changes in temperature and
pressure and has the ability
to readily diffuse into
other gases.
BOILING AND FREEZING
POINTS
Water has unusually high
boiling and freezing point
temperatures, compared to
other compounds with similar
molecular structure. All
other compounds with similar
molecular structure are
gases at ordinary
temperatures. However, due
to the polar nature of the
water molecule, and hydrogen
bonding, the boiling point
of water is a remarkable
100° C and the freezing
point is a remarkable 0° C.
The boiling point of the
most similar substance
(Hydrogen Sulfide) is –60° C
and the freezing point is
–84° C.
FREEZING
Generally, substances
contract (become denser)
with a decrease in
temperature, and water is no
exception. Between 4° C and
the freezing point at 0° C,
an amazing thing happens,
water gradually expands
becoming less dense. Since
the density of ice is less
than that of liquid water,
ice floats on water. About
one-eleventh of the liquid
volume is added at freezing.
It is very significant that
ice expands and floats on
water. The consequences of
this action can be seen in
broken water lines in the
winter and potholes in the
roads. In fact, the freezing
and thawing action of water
is largely responsible for
the fracturing of rock and
the formation of soils.
Also, consider the
consequences if lakes and
streams froze from the
bottom to the top – aquatic
life would not even exist,
and climate and weather
patterns would be altered
drastically.
HEAT CAPACITY
Another remarkable property
of water is its extremely
high capacity to absorb
heat, without a significant
increase in temperature. For
example, the summer sun at
the beach will increase the
temperature of the sand to
the point that it is too hot
to walk on; however, the
water is cool to the touch.
Both the sand and the water
absorb the same amount of
heat energy, but the
temperature of the sand is
higher than the water
temperature. The high heat
capacity of water makes it a
good coolant to use in
condensers and automobile
radiators to keep engines
from over-heating. Water can
absorb about five times the
amount of heat of sand for
an equivalent increase in
temperature.
The moderate climate in
coastal areas is the result
of the absorbing of huge
amounts of solar heat energy
by water during the day and
the slow release of heat
energy during the night.
Inland areas, away from the
coast, typically experience
much wider temperature
extremes. The vast oceans on
earth (about 75 percent of
the surface area) are
responsible for tempering
the climate on earth,
permitting life to exist.
UNIVERSAL SOLVENT
A solvent is a substance
capable of dissolving
another substance (solute)
to form a homogeneous
mixture (solution) at the
molecular level. The highly
polar nature of water makes
it an excellent solvent,
especially for other polar
compounds – salts, alcohols,
carboxyl compounds, etc.
More substances dissolve in
water than in any other
solvent. More than half of
the known elements can be
found in water, some in high
concentrations, and others
only in trace amounts. The
ability of water to dissolve
a substance depends on the
chemical composition,
chemical bond strengths of
the elements, temperature,
and pH. Non-polar compounds
including most hydrocarbons
are difficult to dissolve in
water and dissolve in low or
trace amounts. For example,
oils tend to float on the
surface of water.
SURFACE TENSION
Hydrogen bonding is
attributed to the ability of
water to adhere to or “wet”
most surfaces; such
substances are said to be
hydrophilic (water-loving).
Other substances such as
oils, fats, waxes, and many
synthetics (polypropylene,
etc.) will not “wet” with
water; these substances are
hydrophobic (water-fearing).
Membrane filter cartridges
for water filtration, made
from hydrophobic polymers
and with sub-micron pore
sizes (< 1 micron), must be
manufactured with wetting
agents to lower the surface
tension of the water to
allow the water to “wet” the
pores of the filter. Once
the pores are filled with
water, water will stay in
the pores due to the surface
tension; this is termed
capillary action. Capillary
action is responsible for
the movement of water
through soils, blood through
blood vessels, and water
carrying nutrients through
the roots of plants.
MEDIUM OF LIFE
Water is an essential
ingredient for the existence
of life, as we know it. This
explains the recent interest
in discovering water in
other parts of the universe.
All known biochemical
processes occur in aqueous
environments. The
composition of most living
things is 70 to 80 percent
water, by weight.
Water plays a significant
role in the process of
photosynthesis. In
photosynthesis plants
utilize radiant energy from
the sun to convert two
inorganic substances, water
and carbon dioxide, into
carbohydrates and oxygen.
Photosynthesis is the most
basic and significant
chemical reaction on earth.
It supplies the primary
nutrients, directly or
indirectly, for all living
organisms and is the primary
source of atmospheric
oxygen.
CONCLUSION
To the average person, water
is an ordinary substance,
often taken for granted.
Even though the cause of
these unique and unusual
properties is explainable,
at the atomic level, water
is truly a remarkable
substance. From our
examination of these
properties it is evident
that water is essential for
life, as we know it, to
exist on earth. Water is the
mediator of chemical and
biochemical processes. Water
shapes our natural
environment and even
mediates our climate and
weather. Without water we
would not be here.
Charles Riley, Jr. PE
What is Reagent
Grade Water?
Reagent grade water is water
suitable for use in methods
of analysis and testing
which will not interfere
with the specificity,
accuracy, and precision of
the procedure. Emphasis is
on the specific chemistry of
the procedure and the use of
the water rather than how
the water is prepared.
What are applications for
reagent grade water?
* Proper glassware
washing and rinsing is
critical to prevent
contamination which could
interfere with testing
procedures. This is a common
problem in some critical
applications with
insufficient quality
control.
In instrumental analytical
methods, pure water is used
for blanks, preparing
standards, and sample
dilutions.
Who establishes
reagent grade water
specifications?
Based on the definition of
reagent grade water, the
user of the water should
establish the appropriate
specifications for the
intended application.
However, guidelines have
been established for more
routine applications by such
organizations as the
American Chemical Society (ACS),
College of American
Pathologists (CAP), National
Committee for Clinical
Laboratory Standards (NCCLS),
and the American Society for
Testing and Materials
(ASTM).
Specifications:
Type I water must be
produced on demand without
storage to meet resistivity
specifications. Methods of
production include
deionization, distillation
with polishing deionization,
or reverse osmosis with
polishing deionization.
The specified method of
testing the resistivity of
Type I water is a
temperature compensated
(25°C), in-line monitor.The
specified method of testing
for silicate is a
spectrometric method based
on the reduction of
silicate-molybdate complex.
The pH of Type I and Type II
water cannot be measured
electrometrically because
the water is unstable; the
water is devoid of dissolved
solids and therefore
unbuffered. The slightest
amount of carbon dioxide or
ammonia will radically shift
measured pH.
Type II water is prepared by
distillation or
deionization. Type III water
is prepared by deionization
or reverse osmosis.
Type I water must be
produced on demand and
cannot be stored.
Type II water must be
prepared by distillation
with the appropriate
pretreatment; this water is
generally sterile, pyrogen-free,
storage recontaminates the
water.
What is an
integrated reagent grade
water system?
An integrated water system
is a system that takes
advantage of the
effectiveness of the various
treatment technologies for
specific contaminates and
integrates these
technologies in a synergetic
manner to achieve a very
high level of overall
performance.
System walk-through:
-
Pretreatment with
reverse osmosis improves
the overall water
quality as well as the
economy of operation.
-
Complete recirculation
helps to control
bacteria growth and
provides consistent
quality.
-
Primary treatment with
mixed-bed deionizers
provide ultimate
demineralization; resin
selection is an
important consideration
to maintain quality (use
Type I anion to remove
weakly ionized species
like silica and to
reduce TOC extractables).
-
Ultrafiltration removes
microorganisms, pyrogens,
and colloids.
-
Ultraviolet (254nm) at
the point-of-use
provides sterile water
and prevents retrograde
growth.
-
Ultraviolet (185nm) in
the recirculation loop
reduces TOC to less than
10 PPB.
What is water system
validation?
Validation is the process of
establishing documented
evidence that provides a
high degree of assurance
that the system consistently
produces water meeting the
established specifications.
Written protocols establish
installation and operational
qualifications; process
performance qualifications;
sampling locations and
frequency; testing methods;
instrumentation calibration,
etc.
Charles Riley, Jr.
PE
Introduction
Water is an exceptionally
aggressive solvent that
attacks most of the
substances it contacts. More
substances dissolve in water
than any other solvent. Most
of the known elements can be
found dissolved in water,
some in high concentrations
and others only in trace
amounts. As water moves
through the natural
hydrologic cycle, it
dissolves substances it
contacts. Contaminants
include atmospheric gases
(oxygen, nitrogen, and
carbon dioxide), dissolved
minerals and organic
substances, and suspended
colloidal matter. Water also
provides an ideal
environment for the growth
of bacteria and other
microorganisms if the
necessary nutrients and
conditions for growth exist.
Depending on the type and
concentration of
contaminants, most natural
waters are not suitable for
potable use much less for
most research and industrial
applications. Most all
municipalities and other
purveyors of potable water
provide some level of water
treatment to make the water
suitable for consumption.
The U.S. Environmental
Protection Agency has
established legally
enforceable National Primary
Drinking Water Regulations (NPDWR)
for public water systems.
These regulations are
published on the U.S. EPA
web site (www.epa.gov).
Most high-purity water
systems provide additional
treatment to potable water
to remove residual
contaminants to meet the
water quality specifications
for the given application.
Reagent grade water (RGW)
specifications have been
established by such
organizations as the
National Committee for
Clinical Laboratory
Standards (NCCLS) and the
American Society for Testing
and Materials (ASTM). The US
Pharmacopoeia (USP)
establishes specifications
for compendial water used in
the manufacturing of drug
products. The two major
compendial water types are
USP purified water and USP
water for injection. Many
industries have established
unique water quality
standards specific for their
use.
Contaminants in
Water
To design a high purity
water system, the specific
contaminants in the source
water must be identified and
measured. The NCCLS
classifies water
contaminants based on six
general categories:
dissolved solids (inorganics),
dissolved organics,
dissolved gases, particulate
matter, microorganisms, and
endotoxins (bacterial
by-products that are toxic
in injectable drug
products).
Dissolved Solids
Total dissolved solids (TDS)
include all the ionized
inorganic salts in solution.
Dissolved salts ionize into
their respective cations and
anions in water and
contribute to electrical
conductivity. The
approximate TDS
concentration in water can
be determined by measuring
the electrical conductivity
or resistivity. In pure
water, a relatively small
number of water molecules
ionize into hydrogen and
hydroxyl ions; therefore,
pure water is a relatively
poor conductor of electrical
current. The theoretical
resistivity of pure water is
about 18.2 megohm-cm
(18,200,000 ohm-cm). In
contrast, most potable
waters have resistivities
ranging from about 10,000
ohm-cm to 1,000 ohm-cm. The
relationships between
resistivity, conductivity,
and approximate TDS are
shown in the following
table.
It should be noted that
resistivity in units of
megohm-cm is the reciprocal
of conductivity in units of
microSiemens/cm (?S/cm).
Common cations (positive
charged ions) in potable
water include sodium (Na+),
potassium (K+), calcium
(Ca++), magnesium (Mg ++),
ferrous iron (Fe++), and
aluminum (Al+++). Trace
amounts of heavy metals such
as lead, zinc, and copper
can also be present. Total
hardness includes the
calcium and magnesium salts;
the main concern with
hardness is scale formation
in hot water heaters,
distribution piping, stills,
and reverse osmosis
membranes. Ion-exchange
softening is generally used
for treatment of water to
boilers, stills, and reverse
osmosis systems to prevent
hardness scaling. Softening
removes most of the cations
in water replacing them with
sodium ions in a selective
ion-exchange process.
At high pH ranges (> 7.0 pH)
and in the presence of
oxidizing agents such as
dissolved oxygen or
chlorine, ferrous iron
readily oxidizes to
insoluble ferric iron
(Fe+++) which can foul
ion-exchange resins and RO
membranes. Treatment methods
for removal of iron include
manganese greensand and
Pyrolox. These manganese
oxide medias convert soluble
iron and manganese to
insoluble oxides, then
filters them from the water.
Softening can also be
effective in some
circumstances.
Common anions (negative
charged ions) in potable
water include chloride (Cl-),
sulfate (SO--), nitrate
(NO3-), carbonate (CO3--),
bicarbonate (HCO3-), etc.
Total alkalinity includes
carbonate, bicarbonate, and
hydroxyl ions. If present
with hardness cations and
some heavy metals,
alkalinity contributes to
scale formation.
Silica (SiO2) is one of the
most common elements on
earth, and it is very common
in natural waters. Silica
concentrations in water can
create scaling problem in
boilers, stills, reverse
osmosis membranes, and
cooling water systems.
Silica is usually present in
two forms: ionic silica
(reactive) as SiO2 complex,
and colloidal (non-reactive)
particles. Ionic silica is
weakly ionized, but it can
be removed by ion-exchange,
reverse osmosis, or
distillation. Colloidal
silica is present in
sub-micron particles (less
than 0.1 microns) and can be
removed by ultrafiltration,
reverse osmosis, or
distillation.
TDS removal is accomplished
using such technologies as
ion-exchange
demineralization,
electrodeionization (EDI),
reverse osmosis, or
distillation.
Dissolved Organics
Dissolved organics in water
occur from the natural
degradation of vegetation
and animal wastes and as
pollution from synthetic
compounds such as pesticides
and other chemicals in
industrial discharges.
Naturally occurring
contaminants include
compounds such as tannins (humic
and fulvic acids). These
compounds cause color in
water and can foul
ion-exchange resins and RO
membranes. Free chlorine can
react with some of these
compounds and form
trihalomethanes (THM) which
are determined to be
carcenogenic (cancer
causing).
Many synthetic organic
compounds are used in
industry and agriculture and
can be found in natural
water from industrial
discharges or by leaching
and runoff from soils. Many
of these compounds have
significant health
consequences and are
regulated by the EPA.
Total organic carbon (TOC)
is a direct, quantitative
measure of the amount of
oxidizable, carbon-based
organic matter in water. TOC
concentrations in potable
water typically range from
about 5 to 20 PPM.
Adsorption with activated
carbon is the most effective
means for removing most
dissolved organic
contaminants. Reverse
osmosis is also effective
for removing organic
compounds larger than about
150 molecular weight (MW) in
size; TOC rejections of
greater than 95 percent can
be achieved with RO
membranes. The use of 185 nm
ultraviolet is an effective
method for further reducing
trace dissolved organic
contamination in high purity
water.
Dissolved Gases
Dissolved gases include
carbon dioxide, oxygen,
nitrogen, and hydrogen
sulfide. Carbon dioxide
(CO2) is moderately soluble
in water and can absorb in
water from the atmosphere;
however, most of the carbon
dioxide in natural waters
comes from the carbonates
dissolved in water. Carbon
dioxide reacts with water
forming carbonic acid.
Carbonic acid weakly ionizes
forming bicarbonates and
carbonates. The distribution
of carbon dioxide,
bicarbonates, and carbonates
is a function of the pH of
the water. Carbon dioxide
reduces the pH of water and
is responsible for corrosion
in water lines and boilers.
Although weakly ionized,
carbon dioxide can be
removed by ion-exchange and
de-aeration methods.
The solubility of
atmospheric gases in water
is directly proportional to
the partial pressure of the
gas above the solution; this
is known as Henry’s Law.
Accordingly, oxygen and
nitrogen gas dissolve in
water to saturation levels.
Temperature and TDS content
can also affect solubility.
Nitrogen is an inert gas,
but dissolved oxygen is a
strong oxidizing agent.
Dissolved oxygen is
responsible for corrosion in
water lines, boilers, and
heat exchangers. Dissolved
gases can be removed using
various de-aeration methods
including vacuum degasifiers
and gas transfer membrane
contactors. Oxygen can also
be removed by an
ion-exchange process or by
chemical scavenger agents.
Chlorine gas (added for
disinfection) reacts with
water to form hypochlorite
ion (ClO-) and hypochlorous
acid (HClO); the relative
amounts of each depend on
the pH of the water.
Hypochlorus acid is the more
effective disinfectant and
it is formed at pH values
less than 7.0. Free chlorine
is also known to react with
residual organic compounds
to form trihalomethanes (THMs).
Many purveyors of potable
water are now adding ammonia
gas with chlorine to form
monochloramines (NH2CL).
Chloramines are not as
effective as free chlorine
for disinfection, but they
minimize THM formation and
are more stable and longer
lasting than free chlorine.
Chlorine and chloramines are
effectively removed with
granular activated carbon or
by injection of chemical
reducing agents such as
sodium bisulfite.
Hydrogen sulfide (H2S) is
primarily found in well
water supplies where
anaerobic conditions or
bacterial action reduce
sulfate to sulfide. Hydrogen
sulfide has a characteristic
rotten egg smell. Various
oxidation methods such as
chlorine, ozone, or
oxidizing filters can remove
hydrogen sulfide.
Particulate Matter
Suspended particulate matter
in water may be inorganic or
organic and includes
colloidal silica, fine silt,
organic acids,
microorganisms, and other
discrete dispersed matter.
Generally, the small size
prevents rapid settling, or
the particles are dispersed
by electrostatic surface
charges. Turbidity is the
term used to define this
type of contamination.
Turbidity is measured using
light scattering optical
methods such as a
Nephelometer. Turbidity is
not an absolute measure of
the concentration of
particles, but it is a
relative measure based on
standard stabilized
solutions of various
suspensions such as formazin.
Turbidity is removed using
various filtration methods
including multimedia
filters, ultrafilters,
membrane filters, and
reverse osmosis membranes.
Silt Density Index (SDI) is
another measure of
particulate matter, and the
measured value reflects the
rate of plugging of a 0.45
micron membrane filter disc
by particles in the source
water. The test is used to
correlate the level of
suspended solids in water
that tends to foul reverse
osmosis membranes. Most RO
membrane manufacturers
specify SDI values of less
than 3.0 in the feed water.
Microorganisms
Most bacteria found in
purified water systems are
Pseudomonas species. These
bacteria are generally plant
pathogens found in soil and
water, but a few species are
known to be human pathogens.
They are highly motile
(flagellated), live in an
aerobic environment, and
oxidize glucose for
nutrients (heterotrophic).
They are rod shaped single
cell microorganisms about
0.5 microns in diameter and
about 3 to 5 microns in
length. They are
opportunistic and can adapt
and survive under severe
conditions of extremely low
concentrations of organic
substrates such as in
purified water systems. The
slimy polysaccharide cell
wall of the bacteria
promotes adhesion to
surfaces and biofilming
occurs rapidly on any
contact surface. The
polysaccharide coating also
traps nutrients and protects
the cell from disinfectants
such as chlorine.
Bacteria are quantified in
terms of colony forming
units per volume of water (CFU/ml
or CFU/100 ml). Bacteria
testing involves filtering a
known volume of water
through a sterile 0.45
micron membrane filter disc,
incubating the filter disc
on a nutrient pad at a
standard temperature (35oC),
and counting the colonies
formed on the filter disc
after the prescribed
incubation time (48 hours).
Each colony is assumed to
have grown from a single
cell. This is termed the
“standard plate count”
method.
Since bacteria replicate
rapidly under ideal
conditions for growth,
control of bacteria in a
purified water system is one
of the most difficult
challenges. The best
strategies include the
following: provide complete
recirculation at turbulent
velocities (3 fps); use 254
nm ultraviolet in the loop;
eliminate any piping
dead-legs; use sanitary
piping with low surface
roughness; operate at
sanitizing temperatures; or
frequently flush and
sanitize with hot water,
ozone, or other chemical
disinfectants. Even with
these measures, biofilming
can occur in ambient
temperature systems and
compromise the
bacteriological quality of
the water. An inexpensive,
yet effective measure for
controlling bacteria at the
point of use is to install
sterile 0.2 micron membrane
filter capsules on the water
faucets or outlets.
Endotoxins
Endotoxins are the
polysaccharide compounds
from the cell wall of
certain bacteria such as
those found in purified
water systems. They are
termed pyrogenic because
they induce a fever response
when injected in
warm-blooded mammals and can
even cause shock and death.
They have two major
components: a hydrophilic
(water soluble)
polysaccharide chain
attached to a hydrophobic
(insoluble in water) lipid
(fatty) group. The
hydrophobic portion causes
endotoxins to aggregate
together in vesicles ranging
in size from about 20,000
daltons to millions of
daltons. One molecular
weight (MW) is about one
dalton. Based on this,
endotoxins can be removed
using ultrafilter membranes
in the 10,000 dalton pore
size range or by using
reverse osmosis membranes.
Properly designed
distillation systems can
also remove endotoxins.
Endotoxins are quantified in
terms of Endotoxin Units per
milliliter (EU/ml) using the
Limulus Ameobocyte Lysate
test (LAL). EU’s are
assigned by comparison with
a USP reference endotoxin
standard. Endotoxins react
with the LAL (purified
extract of the blood of the
horseshoe crab) causing a
turbid or clotting reaction
that permits quantification
to extremely low levels
(about 0.001 EU/ml). The USP
water for injection
specification limit for
endotoxin is less than 0.25
EU/ml.
Water Quality
Specifications
Reagent Grade Water (RGW)
RGW is defined as water
suitable for use in a
specified procedure such
that it does not interfere
with the specificity,
accuracy, and precision of
the procedure. In addition,
the water quality must meet
the specifications
established for the
application. This definition
applies to any high purity
water application.
CAP, NCCLS, and ASTM have
established RGW
specifications for uses
ranging from general
laboratory to specific
clinical laboratory
applications. General
laboratory applications
include glassware washing
and rinsing, chemical
reagent and buffer solution
preparation, making blanks
and standards for
calibrating analytical
instrumentation, culture
media, etc. Clinical
laboratory applications
include procedures in
bacteriology, immunology,
hematology, histology, etc.
The NCCLS reagent grade
water specifications are
shown in Table 1. The ASTM
reagent grade water
specifications are shown in
Table 2.
Some applications may have
“special” requirements
beyond RGW specifications.
For example, high
performance liquid
chromatography (HPLC) may
require water with a maximum
absorbence of a specified
wavelength of ultraviolet
light. Special “HPLC” grade
water systems are offered by
some companies, and some
suppliers offer “HPLC” grade
bottled water.
National Committee for
Clinical Laboratory
Standards (NCCLS)
The NCCLS specifies three
grades of RGW (Types I, II,
and III). The NCCLS does not
specify the acceptable
methods of water
purification for producing
RGW; however, it does state
that any method or
combination of methods is
acceptable as long as the
product water meets the
applicable specifications.
Type I water is the highest
quality and is generally
used in more critical
applications such as trace
element analysis, automated
analyzer systems, reagent
and buffer solution
preparation, etc. Type II
water is used in general
clinical methods including
immunology, hematology, etc.
Type III water is used for
some qualitative procedures,
glassware washing, etc.
The NCCLS specifies water
sampling and testing methods
for the parameters listed.
Type I water quality must be
measured using an inline
resistivity sensor to avoid
the problems associated with
rapid absorption of
atmospheric carbon dioxide
into the deionized water.
Rapid absorption of even
small amounts of carbon
dioxide into the water
sample causes a significant
drop in the resistivity of
the water.
In addition, general water
purification system design
and maintenance guidelines
are offered by NCCLS. It is
suggested to use inert
materials of construction to
prevent leaching of
inorganic and organic
contaminants. Systems should
be designed with complete
recirculation avoiding
dead-legs (stagnant areas).
Outlet designs should
minimize dead spaces and use
non-leaching seal materials.
It is generally not
preferred to store and
distribute pure water (Type
I water can not be stored
and remain Type I quality);
it is suggested to produce
final product water quality
on demand at the
point-of-use. Systems
designed to store and
distribute Types II and III
water should be provided
with measures to protect the
chemical and microbial water
quality (recirculation with
254 nm ultraviolet, sealed
tanks with 0.2 micron
hydrophobic vent filters,
etc.). Sanitization of the
system is recommended at
least semi-annually or as
necessary for quality
control.
NCCLS, Type I water systems
must include granular
activated carbon treatment
for organics and chlorine
removal, mixed-bed
deionization to meet
resistivity and silica
specifications, and 0.2
micron post-filtration for
bacteria and particle
control. Type II water can
generally be produced by
distillation, deionization,
or reverse osmosis with
polishing deionization or
electrodeionization (EDI).
Reverse osmosis technology
is capable of providing Type
III reagent grade water
depending on the feed water
quality and the design and
operation of the reverse
osmosis system.
Table 1
National Committee for
Clinical Laboratory
Standards
Reagent Grade Water
Specifications
|
Parameter |
|
Type II |
Type III |
|
Bacteria, max. (CFU/ml) |
|
|
|
|
pH, units |
|
|
|
|
Resistivity, min. (megohm) |
|
|
|
|
Silica, max. (mg/l) |
|
|
|
|
Particles |
|
|
|
|
Organics |
|
|
|
The American Society for
Testing and Materials (ASTM)
The ASTM establishes
specifications for Types I,
II, III, and IV reagent
grade water (D1193-99e1) as
shown in Table 2. In
addition, the water quality
is further classified as
Type A, Type B, or Type C
depending on the applicable
bacteriological and
endotoxin quality. Type I
water is the highest quality
and is generally used for
the most critical
applications – trace element
analysis, HPLC, reagent
preparation, etc. The ASTM
further specifies that Type
I water is produced by
mixed-bed deionization with
suitable pretreatment
(distillation or other equal
process that can produce
water with a maximum
conductivity of 20 uS/cm)
and post filtration with 0.2
micron membrane filters.
Type I water quality can not
be maintained in storage and
must be produced on demand
at the point-of-use.
Resistivity can only be
measured using inline
resistivity monitoring
equipment.
Type II reagent grade water
is produced by distillation
with suitable pretreatment
(reverse osmosis or
deionization) and, depending
on the design of the storage
tank, is generally sterile
and endotoxin-free. This
grade of water is suitable
for preparing culture media,
microbiology, bacteriology,
etc. Care must be taken in
the design of the storage
tank and the distribution
system to prevent bacterial
contamination.
Type III reagent grade water
is produced by distillation,
deionization, reverse
osmosis, electrodeionization,
or a combination of these
technologies, followed by
post-filtration with a 0.45
micron membrane filter. This
grade of water is generally
suitable for preparing
various reagents,
qualitative analysis, etc.
Design of storage tanks and
distribution systems is
critical to prevent
contamination.
Type IV reagent grade water
is produced by any of the
primary treatment methods
(distillation, deionization,
electrodialysis, or reverse
osmosis) or a combination of
these methods. This water
quality is generally used
for glassware washing,
cooling applications, etc.
Table 2
American Society for Testing
and Materials
Reagent Grade Water
Specifications
|
Parameter |
Type I |
Type II |
Type III |
Type IV |
|
Resisitivity, min. (megohm) |
18.0 |
1.0 |
4.0 |
0.2 |
|
pH, units (25°C) |
NA |
NA |
NA |
5 - 8 |
|
TOC, max. (ug/l) |
50 |
50 |
200 |
NS |
|
Sodium, max. (ug/l) |
1 |
5 |
10 |
50 |
|
Chloride, max. (ug/l) |
1 |
5 |
10 |
50 |
|
Total Silica, max. (ug/l) |
3 |
3 |
500 |
NA |
|
|
Type A |
Type B |
Type C |
|
Bacteria, max. (CFU/100
ml) |
1 |
10 |
1000 |
|
Endotoxin (EU/ml) |
<0.03 |
0.25 |
NA |
Pharmaceutical Grade Water
The United States
Pharmacopoeia (USP 25) is
the accepted guide
(compendium) for producing
pharmaceutical products in
the US. The USP specifies
standards of quality,
purity, packaging, and
labeling for many
pharmaceutical products and
the ingredients used in the
manufacture of these
products. The guide
specifies two grades of bulk
ingredient water used in the
preparation of compendial
dosage forms, USP purified
water (PW) and USP water for
injection (WFI).
The US Food and Drug
Administration (FDA)
provides a regulatory
/enforcement function in the
pharmaceutical industry. The
FDA has established
guidelines known as the
Current Good Manufacturing
Practices (CGMP) to regulate
the industry. The CGMPs are
essentially guidelines used
by industry to establish
comprehensive quality
management plans to insure
the safety of drug products.
The CGMPs provide some
guidelines that affect the
design and operation of USP
grade water systems. In
addition, water systems must
be validated systems;
validation is a documented
process that provides
assurance that the system
will remain in control and
consistently provide the
specified water quality.
USP Purified Water (PW)
USP purified water is an
ingredient in many
pharmaceutical products. It
is not applicable for the
manufacture of parenteral (injectable)
drugs and some other drug
products. The USP specifies
the following requirements
for purified water:
• Produced by a suitable
process – reverse osmosis,
deionization, or
distillation
• Produced from “drinking
water” – meets EPA potable
water standards
• Contains no added
substances
• Meets the requirements of
the Conductivity Test – less
than 2.1 ?S/cm at 25oC, and
6.6 pH (varies depending on
temperature and pH)
• Meets the requirements of
the TOC Test – less than 500
PPB
• To control bacteria
growth, USP suggests a
microbial action limit of
100 CFU/ml
The FDA stipulates the
following additional
requirements for purified
water:
• Free of “objectionable
organisms” that can infect
or grow in the product – the
burden is on the
manufacturer to determine
the nature of this problem
and establish applicable
specifications
• Conformance with CGMPs
USP Water for Injection
(WFI)
USP water for injection (WFI)
and sterile water for
injection are used in the
manufacture of parenteral
drugs and other products
such as ophthalmic and
inhalation products. The USP
specifies the following
requirements for water for
injection:
• Meets all of the
requirements for purified
water
• Produced by distillation
or reverse osmosis
• Meets the requirements of
the Bacterial Endotoxin Test
– less than 0.25 EU/ml
• Suggested microbial action
limit of 10 CFU/100ml
• Produced, stored, and
distributed under conditions
designed to prevent
production of endotoxins
The FDA stipulates the
following additional
requirements for WFI:
• Reverse osmosis as a final
means of production is
discouraged, but if it is
used, double-pass reverse
osmosis is suggested
• Conformance with CGMPs
CGMP Compliance Issues
Critical process parameters
are those parameters that
directly affect water
quality and include
conductivity, TOC,
sanitizing temperature, etc.
Because microbial quality
and endotoxin quality cannot
be monitored in real time,
monitoring of parameters
that affect these are
considered critical and may
include process temperature,
UV intensity, ozone
concentration, recirculation
flow rate, loop pressure,
etc. Instruments that
measure these parameters are
considered critical
instruments, and these
instruments must be properly
calibrated on a regular
basis and documentation must
be maintained.
Regulatory compliance issues
have more to do with
establishing system design
specifications and properly
documenting the validation
of the system. This includes
verifying installation,
operational, and performance
qualifications; establishing
sampling and testing
protocols; materials
verification documentation;
establishing sanitization
protocols; record keeping;
etc.
The critical concern in the
design of pharmaceutical
water systems is to avoid
designs that could
compromise bacterial and
endotoxin quality. CGMP
design strategies to control
bacterial and endotoxin
quality include the
following:
-
Provide continuous
recirculation at
turbulent flows (3 fps
in the return line)
-
Eliminate piping
dead-legs and use zero
dead-leg outlet valves
-
Establish standard
operating procedures
(SOP) for frequent
draining, flushing, and
sanitizing
-
Maintain positive system
pressure
-
Adhere to accepted back
flow prevention practice
-
Design for complete
system drain ability (WFI
systems)
-
Provide smooth, inert
surfaces and joints (180
grit minimum)
-
Store and distribute WFI
quality water at 80o to
90oC
General Design
Considerations
It is important to establish
the appropriate water
quality specifications for
the given application. If
specific regulatory or
industry standards do not
exist, the user is
responsible for establishing
the applicable water quality
specifications.
It is generally impractical
to design individual water
systems to provide specific
water quality for each
application in a large
facility. If the most
critical application
requires the bulk of the
water usage, it is common to
design a system to meet
these requirements and
supply this quality of water
to all applications.
However, if the critical
application requires minimal
water usage in comparison to
other less critical
applications, providing
point-of-use polishing
systems to meet the needs of
the most critical
applications may be the most
practical option. Economic
considerations generally
influence the system design.
The design of the water
system may also be
influenced by the applicable
water quality
specifications. For example,
ASTM and USP specify the
method of production for the
various types of water
quality.
A water system designed to
provide ASTM, Type I reagent
grade water for a small
laboratory consists of a
number of integrated
components. A pre-filter
cartridge (0.2 to 0.5 micron
pore size rating) provides
for removal of particulate
matter in the feed water. A
granular activated carbon
filter removes chlorine and
organics. A primary
mixed-bed deionizer provides
TDS removal to a maximum of
20 uS/cm conductivity, and a
polishing mixed-bed
deionizer provides final
product water quality with a
conductivity of 0.055 uS/cm.
In addition, the system may
include an in-line
conductivity sensor and
monitor for final water
quality, a dispensing faucet
with a 0.2 micron
post-filter capsule, and a
recirculation pump (Figure
1). A sterile 0.2 micron,
absolute rated membrane
filter cartridge can produce
bacteria-free water.
Optional components can
include a 185 nm ultraviolet
system installed before the
primary deionizer to provide
for trace organics removal
(TOC less than 10 PPB), an
ultrafilter membrane for
endotoxin removal, and a 254
nm ultraviolet unit for
bacteria control.
For greatly improved water
quality and economy of
operation (especially for
higher capacity systems), a
reverse osmosis system can
be used in lieu of the
primary deionizer unit.
Reverse osmosis is a
membrane separation process
that filters water through a
membrane with pore sizes
less than 0.001 micron and
can provide greater than 98
percent rejection of TDS in
the feed water. In addition,
reverse osmosis systems
reject a high percentage of
organics, colloids,
bacteria, and pyrogens. The
throughput of the polishing
mixed-bed deionizers can be
increased by as much as 6 to
10 times or more (depending
on the feed water quality)
by using reverse osmosis for
primary water treatment.
Reverse osmosis membranes do
not reject dissolved carbon
dioxide in the feed water,
and this is an additional
load on the polishing
mixed-bed deionizer. For
many small laboratory
systems the capital cost
payback can be as short as a
few months.
Large central systems
designed to provide ASTM,
Type I reagent grade water
generally include a reverse
osmosis system for primary
treatment, storage tank with
water distribution pump
station, polishing mixed-bed
deionization, 0.2 micron
post-filter, and a
recirculated piping system
(Figure 2). An optional 254
nm ultraviolet system may be
installed after the
post-filter for better
control of bacteriological
quality.
Applicable pretreatment for
reverse osmosis systems
depends on the feed water
quality and may include a
turbidity filter, carbon
filter for chlorine removal,
softening system for
hardness removal, and a 5
micron disposable cartridge
pre-filter Pretreatment is
critical to optimize the
performance and the life of
the reverse osmosis
membranes.
One of the most difficult
problems in the design of
large central water systems
to provide ASTM, Type I
water quality is maintaining
Type I water quality at
every point-of-use location
in a complex water
distribution system. The
design of the water
distribution system is
critical to maintaining
water quality. Low velocity
or stagnant areas in the
piping system, leaching of
contaminants from the piping
materials, bacteriological
activity, and intrusion of
carbon dioxide in the water
can rapidly degrade water
quality.
Another common design is to
provide a central system
producing ASTM, Type III
water quality or NCCLS, Type
II water quality and install
polishing mixed-bed
deionization systems at
critical points-of-use where
Type I water quality is
required.
ASTM, Type III and Type IV
reagent grade water must be
produced by distillation,
mixed-bed deionization, or
electrodeionization (EDI) to
meet the strict sodium and
chloride specifications. A
typical system may consist
of reverse osmosis for
primary treatment (with
applicable pretreatment
equipment) followed by
mixed-bed deionization or
EDI, water storage tank,
booster pump station, and a
recirculated piping system.
To consistently meet the
resistivity specification
for ASTM, Type III water,
polishing deionization may
be required after the
storage tank. Alternatively,
a nitrogen blanketing system
on the storage tank to
prevent the absorption of
carbon dioxide may be
required to prevent
degradation of the
resistivity of the stored
water. Applications with
minimal water usage
requirements are often
served with exchangeable
mixed-bed deionizers, even
though the water quality may
be better than needed.
Other integrated system
design options can be used,
but a thorough knowledge of
the feed water quality,
water quality objectives,
and water system process
design is needed to provide
a suitable system.
Common practice in high
purity water system design
for pharmaceutical and
biotechnology applications
is to store the final
product water (PW or WFI)
before distribution to the
points-of-use. This practice
facilitates frequent
sanitization of the system
or allows for storage and
distribution of the water at
sanitizing temperatures
(90oC) which is common in
WFI systems.
Many combinations of unit
processes can be used to
produce USP purified water.
Final treatment options
include distillation,
mixed-bed deionization,
double-pass reverse osmosis,
reverse osmosis followed by
polishing mixed-bed
deionization, and reverse
osmosis followed by EDI.
The most common USP purified
water system design consists
of an RO system for primary
water treatment (with
applicable pretreatment
equipment), polishing
mixed-bed deionization, 0.2
micron post-filter, 254nm
ultraviolet system, USP
water storage tank, booster
pump station, and
recirculated distribution
piping system (Figure 3).
Other common process design
configurations include the
following:
-
Pretreatment, RO primary
treatment, EDI polishing
treatment, 0.2 micron
post-filter, 254nm
ultraviolet, storage,
and distribution
-
Pretreatment, RO primary
treatment, storage,
booster pump,
distillation (vapor
compression or
multi-effect), storage,
and distribution
In systems designed for hot
water or steam sanitization,
the tank, booster pump, heat
exchanger, and piping system
are constructed of sanitary,
316L stainless steel. The
storage tanks used in these
systems are ASME Code
pressure vessels constructed
of polished, 316L stainless
steel and rated for vacuum
and pressure applications at
temperatures above the
maximum operating
temperature of the system.
These pressure vessels are
also equipped with sanitary
fittings, 0.2 micron
absolute rated hydrophobic
vent filter, pressure/vacuum
rupture disc, internal spray
ball on the recirculation
return line, level and
temperature controls, and
insulation.
Other design considerations
include sizing of equipment
items to meet capacity and
peak usage demands, utility
requirements, space
availability and access,
instrumentation and control
requirements, maintenance
requirements, and future
needs. Budget considerations
may also influence system
design options. In some
cases, the cost of the high
purity water system can be a
significant percentage of
total process system cost.
Other significant costs
could include system
validation, quality control,
and wastewater treatment.
References
1. National Committee for
Clinical Laboratory
Standards, NCCLS C3-A2,
“Preparation and Testing of
Reagent Water in the
Clinical Laboratory;
Approved Guideline”, Third
Edition, Vol. 17 No. 18,
October 1997.
2. American Society for
Testing and Materials, ASTM
D1193-99e1, “Standard
Specification for Reagent
Grade Water”, Annual Book of
ASTM Standards, Volume
11.01, American Society for
Testing and Materials,
Philadelphia, PA (2003).
3. Meltzer, Theodore H.,
High Purity Water
Preparation for the
Semiconductor,
Pharmaceutical, and Power
Industries, Tall Oaks
Publishing, Inc., Littleton,
CO (1993).
4. International Society for
Pharmaceutical Engineers,
Baseline Pharmaceutical
Engineering Guides for New
and Renovated Facilities,
Water and Steam Systems,
Vol. 4, First Edition,
January 2001.
5. Winstead, Martha. Reagent
Water: How, When, and Why?.
The American Society of
Medical Technologists,
Houston, TX (1967).
6. Murphy, Daniel B., and
Viateur Rousseau.
Foundations of College
Chemistry. New York: The
Ronald Press Company (1969).
|